<?xml version="1.0" standalone="yes"?>
<?xml-stylesheet type="text/xsl" href="..\..\XSLTemplates\ErrorsStyleSheet.xsl"?>
<Error>
<ErrorNumber>15320</ErrorNumber>
<ErrorFamily>Autoloader</ErrorFamily>
<ErrorMessage>
ALE: STEP : axis 2 communication error
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>Trendsetter_400/800_Auto</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<OperatorAction>
<DescriptionText><text>
Remove the plate from the unload table and select the PLATE REMOVED button on
the workstation dialog box. If the problem persists, then Creo Service should be
called.
</text></DescriptionText>
</OperatorAction>
<ServiceInformation>
<DescriptionText><text>
This error occurs when there is a communications problem between the firmware
and the stepper controller on the TEE-board.
</text></DescriptionText>
</ServiceInformation>
<Cause>
If the procedure described in Operator Action does not correct the problem, then
the stepper controller must be re-initialized. Do this either by typing TABLE
INIT or (if that does not work) resetting the recorder.
</Cause>
<SeeAlso>
<UnnumberedList>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:33:01 2008 PST
</Updated>
<Links>
</Links>
</Error>